RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Increase in Pressure (1491)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2024
Event Type  malfunction  
Event Description
Philips received a complaint by the customer on the v60 indicating that the device is getting error code 1211 alarm message: high-rate message.It is unknown if there was patient involvement.The investigation is ongoing.
 
Manufacturer Narrative
It was reported there was no patient involvement at the time the issue was discovered.Per good faith effort response, it was confirmed that the reported issue was caused by the biomed customer misconfiguring the alarms.Based on the information provided and service performed, the customer's alleged malfunction was due to user error.There was no malfunction with the device.Bench repair reconfigured the alarms to resolve the reported issue.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18739128
MDR Text Key335756280
Report Number2518422-2024-08506
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053614
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/09/2024
Initial Date FDA Received02/20/2024
Supplement Dates Manufacturer Received04/14/2024
Supplement Dates FDA Received04/16/2024
Date Device Manufactured03/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-