Model Number TJF-Q180V
Device Problems Crack (1135); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and the evaluation found no additional reportable malfunctions other than those mentioned in b5.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed that during the device evaluation, the evis exera ii duodenovideoscope exhibited a gap between the distal end and the z-block and the adhesive around the objective lens had foreign objects.There were no reports of patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 13 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon of "gap between z-block and c-body [distal end of the device]" was presumed to have been due to physical stress such as hitting/dropping distal end or chemical stress such as chemical solution used.However, it was not possible to further specify the cause of the event.The suggested phenomenon of "lens glue has foreign material" was presumed to have been due to a leakage from the device, disallowing proper reprocessing.As there was no information received of clear misuse of the device, there is no labeling advisory that was deemed required or recommended for the user facility.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18740364
MDR Text Key335770202
Report Number9610595-2024-03514
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K143153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberTJF-Q180V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/14/2024
Initial Date FDA Received02/20/2024
Supplement Dates Manufacturer Received02/27/2024
Supplement Dates FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0379-2022
Patient Sequence Number1
-
-