It was reported that a patient underwent an atrial fibrillation ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and post-procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, the medical team would not able to zero the catheter and the force values were incorrect without the catheter being moved.To troubleshoot the cable was replaced without resolution.The catheter was replaced, the issue was resolved, and the procedure was continued.No patient consequences were reported.
|
The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.Visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis of the returned sample revealed reddish material and a hole in the pebax.The magnetic feature was tested, and no errors were observed.A screening test could not be performed due to no temperature was displayed due to an open circuit in the tip area; however, the force values and the vector displayed during magnetic sensor functionality test were observed within specifications.No force issues were observed.The reddish material found inside the pebax area may contribute to the force issue reported.The root cause of the hole in the pebax could be related to the handling of the device during the procedure however this cannot be conclusively determined.A manufacturing record evaluation was performed for finished device number 31151895l, and no internal action related to the complaint was found during the review.The issue reported by the customer was confirmed; it should be noted that product failure is multifactorial.The instructions for use contain the following information that should be considered: to ensure accurate force readings, verify that the force reading is near zero when the catheter is not in contact with tissue.If the force reading is not near zero when the catheter is not in contact with tissue, perform zeroing.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|