• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA PADGETT MODEL S SLIMLINE ELECTRIC DERMATOME; SKIN GRAFT PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA PADGETT MODEL S SLIMLINE ELECTRIC DERMATOME; SKIN GRAFT PRODUCTS Back to Search Results
Catalog Number 3539700
Device Problem Insufficient Information (3190)
Patient Problem Wrinkling (2613)
Event Date 04/06/2019
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
It was reported that during graft sample procedure, a dermatome (id (b)(6)) "tugged" the skin.No additional information was provided.
 
Event Description
N/a.
 
Manufacturer Narrative
Dermatome was returned for evaluation: failure analysis - the head assembly is misadjusted and need to get readjusted.The 2" width clip is wasted and will be exchanged.Additional some small wasted parts need to get replaced.The power supply is missing all 4 rubber feet, the rubber feet will be added.Have to perform final functional and safety test.Root cause - the most likely root cause is use error and/or rough handling over time causing damage to the dermatome.Proper adjustment, proper technique, and proper handling per the dermatome ifu should prevent against "turbulent" grafts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PADGETT MODEL S SLIMLINE ELECTRIC DERMATOME
Type of Device
SKIN GRAFT PRODUCTS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18741858
MDR Text Key335781159
Report Number3004608878-2024-00024
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number3539700
Device Lot NumberS1579
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/11/2019
Initial Date FDA Received02/20/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/19/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-