Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
Physician attempted to use an everflex entrust self expanding stent with a 5fr sheath and a 0.035 guidewire during treatment of an 180mm fibrous lesion in the patient¿s proximal superficial femoral artery (sfa).Moderate vessel tortuosity and little vessel calcification were reported.Artery diameter reported as 8mm.There were no abnormalities reported in relation to anatomy.No embolic protection was used.There was no damage noted to packaging, (i.E.Shelf carton, hoop/tray).No issues were noted when removing the device from the hoop/tray.The ifu was followed and the device was prepped without issue.The vessel was not pre-dilated.The stent was not passed through a previously deployed stent.No resistance was encountered when advancing the device.No excessive force was used.The thumbscrew/lock-pin was checked for securement prior to procedure.The lock pin was removed after insertion into the body.It was reported the stent was unable to fully deploy.Stent catheter handle had to be opened and manual deployment was completed to deliver the stent.No deformation noted to the deployed stent.The delivery system was safely removed.Procedure was completed and patient did fine.No vessel damage.No patient injury reported.
|