MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3: date of event and d5.Operator of device is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that during preventative maintenance, the device was found not alarming and no leds.Board needs to be replaced.No patient or clinical harm reported.
 
Manufacturer Narrative
H6.Investigation codes: updated.Investigation summary: the affected device was returned for evaluation.A visual inspection was performed, and the led lights were found to be broken.The customer's indicated failure was confirmed as the device alarmed and the led lights were damaged.The root cause was the damaged pcb.No action was taken due to the demand, condition, and age of the device.It was deemed beyond economical repair and will be scrapped.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.D4.Unique identifier (udi) #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18743530
MDR Text Key336742384
Report Number2183161-2024-00152
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public(01)10695085002796(11)191014
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/26/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/22/2024
Initial Date FDA Received02/20/2024
Supplement Dates Manufacturer Received03/11/2024
Supplement Dates FDA Received04/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-