It was reported, the deflectable videoscope, the periphery of the 3d screen becomes blurry.The issue occurred during an the procedure, and the procedure was completed using the same set of equipment.There were no reports of patient harm.
|
This supplemental report is being submitted to provide further information provided by the customer, the results of the legal manufacturer's final investigation and associated h6 coding.Additionally, the following fields were corrected due to inadvertent errors in the initial report: (concomitant medical products and therapy dates), device/therapy date added.Device manufacturer date) added.(investigation findings), code "213 - no device problem found" removed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over one (1) year since the subject device was manufactured.According to the customer, the issue was no longer occurring and therefore, the device will not be returned to olympus.Based on the results of the investigation, as the issue could no longer be produced and the device was not inspected at olympus, root cause cannot be established.It is possible that the user was observing outside the viewable field of the 3d (3-dimensional) monitor.If the monitor is viewed from outside the specified range, the user may not be able to see a normal 3d image.Or one of the two objective lenses needed for 3d viewing was temporarily blocked by contamination (e.G.Body fluid or blood).If one lens view is blocked/stained, the 3d image may not be displayed correctly.Olympus will continue to monitor field performance for this device.
|