|
Catalog Number IAB-06840-U |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/30/2024 |
Event Type
malfunction
|
Event Description
|
It was reported "device alarmed for helium leak.The catheter was not found to be fractured, and the device was tested with a simulator; the device did not alarm.There was blood on the inside of the anterior section of the catheter near the marked spot, presumably the catheter was leaking".The catheter was removed and replaced in the same insertion site to complete the procedure.No patient injury or consequence reported.Patient's current condition reported as "fine".
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
(b)(4).The reported complaint that "device alarmed for helium leak" was confirmed upon investigation of the returned sample.The customer returned a 40cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) without the original packaging for investigation along with the arterial pressure tubing and a driveline tubing; a connector was noted missing from the driveline tubing.Upon return, dried blood was noted within the sheath sidearm and on the exterior surfaces of the returned sample.No blood was noted within the bladder/helium pathway.The intra-aortic balloon catheter (iabc) central lumen was noted damaged within the iabc flex-tip assembly area and a cross-over leak was confirmed; this can result in helium loss alarms.The returned iabc bladder membrane was fully intact, with no leaks noted.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the damaged central lumen.The root cause of the complaint was undetermined.A capa has been initiated under teleflex's quality systems to further investigate the issue.
|
|
Event Description
|
It was reported "device alarmed for helium leak.The catheter was not found to be fractured, and the device was tested with a simulator; the device did not alarm.There was blood on the inside of the anterior section of the catheter near the marked spot, presumably the catheter was leaking".The catheter was removed and replaced in the same insertion site to complete the procedure.No patient injury or consequence reported.Patient's current condition reported as "fine".
|
|
Search Alerts/Recalls
|
|
|