Model Number 8300AB |
Device Problems
Perivalvular Leak (1457); Insufficient Information (3190)
|
Patient Problems
Dyspnea (1816); Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
|
Event Date 01/25/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
Event Description
|
Through implant patient registry it was learned that a 23mm 8300ab valve was explanted after an implant duration of 7 (seven) years, 2 (two) months due to unknown reasons.The explanted valve was replaced with a 25mm 11500a valve.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
|
|
Event Description
|
Through implant patient registry it was learned that a 23mm 8300ab valve was explanted after an implant duration of 7 (seven) years, 2 (two) months due to paravalvular leak.The patient presented with heart failure and shortness of breath.The explanted valve was replaced with a 25mm 11500a valve.The patient was stable post procedure.
|
|
Manufacturer Narrative
|
The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Regurgitation is considered to be a perivalvular leak (pvl) if a turbulent eccentric jet originates between the bioprosthetic sewing ring and the annulus.Pvl can occur in the mitral and aortic position for similar reasons.In the early postoperative period, the highest incidence of pvl has been seen in patients developing infective endocarditis, which is most likely attributed to inadequate peri-operative antibiotic prophylaxis or nosocomial infection.Annular calcification is also a risk factor for the development of peri-operative pvl as the bioprosthesis may not seat properly after debridement.Technique related factors, such as incorrect valve sizing, have been shown to contribute to the development of pvl.Anatomical factors may create difficulty seating the bioprosthetic valve resulting in pvl.The anatomy of the annulus may induce mechanical stresses along the rigid bioprosthetic ring which can influence long-term valve performance and durability.A diseased or rigid annulus can potentially increase the mechanical stress on the prosthetic valve, leading to pvl.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|