Model Number CDHFA500Q
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2024
Event Type  malfunction  
Event Description
It was reported that the patient presented for implant of the implantable cardioverter defibrillator.During the procedure the physician attempted to connect the lead to the left ventricular port.However, the setscrew could not be engaged.A replacement device was used to complete the implant.The patient was stable.
 
Manufacturer Narrative
The reported field event of setscrew connection anomaly was confirmed.Analysis revealed the left ventricular setscrew was stripped and contained septa material inside the hex cavity.This material in the hex cavity prevented full insertion of the torque driver and was the cause of the reported event.The set screw anomaly was consistent with having occurred during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18744417
MDR Text Key335852588
Report Number2017865-2024-33131
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Lot NumberP000185297
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/31/2024
Initial Date FDA Received02/20/2024
Supplement Dates Manufacturer Received03/08/2024
Supplement Dates FDA Received03/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-