ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; NO MATCH Back to Search Results
Model Number 2088TC/52
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/18/2020
Event Type  Injury  
Event Description
It was reported that the right ventricular (rv) lead was capped and replaced on (b)(6) 2020 due to unknown reason.During an unrelated procedure, the capped rv lead was explanted.The patient was in stable condition.
 
Manufacturer Narrative
The lead was returned with insufficient information.A complete lead was returned in one piece.Visual examination of the lead found external abrasions breaching the outer insulation and exposing the outer coil caused by friction to the device can located proximal from the suture tie impressions.Electrical testing of the lead found an internal short between the inner coil and outer coil conductor paths due to procedural damage.Electrical testing of the lead was normal with no anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18744645
MDR Text Key335809126
Report Number2017865-2024-33136
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502887
UDI-Public05414734502887
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2014
Device Model Number2088TC/52
Device Lot Number3425647
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/29/2024
Initial Date FDA Received02/20/2024
Supplement Dates Manufacturer Received05/17/2024
Supplement Dates FDA Received05/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATRIAL LEAD; PACEMAKER
Patient Outcome(s) Required Intervention;
Patient Age48 YR
Patient SexFemale
-
-