• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA FABIUS GS; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA FABIUS GS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8604700
Device Problems Intermittent Continuity (1121); Gas Output Problem (1266); Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/02/2024
Event Type  malfunction  
Manufacturer Narrative
The investigation was solely performed based on the information provided by the dispatched service technician as the device log was not available.During on-site analysis the reported ventilator failure could be confirmed and the unit was found to have lost all information.In consequence the pcb control and the ventilator assembly were replaced.Afterwards, the unit was successfully tested according to manufacturer's specification.As the fabius is 17 years old it can be assumed that wear and tear has caused the failure of single components.In case the fabius shuts down automatic ventilation, the device will generate an audible alarm and a visible alarm message "ventilator fail" will be displayed.In this case automatic ventilation is not possible.The user can switch to manual ventilation as described in the instructions for use.Monitoring is still functional.The device has reacted as specified by posting corresponding alarms to inform the user about the situation.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported that the device experienced a ventilator failure at end of case.No patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FABIUS GS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key18746125
MDR Text Key335848814
Report Number9611500-2024-00066
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K011404
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8604700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/23/2024
Initial Date FDA Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-