It was reported that on (b)(6) 2024, a 21mm sjm regent heart valve was chosen for a procedure.The patient's inr ratio was 2.0~2.5.At the time of preparation, the leaflet was moving normally and it was tested by a leaflet tester.During procedure, it was noted that one of the leaflet was stacked and a gap was observed, so it was rotated and checked the leaflet was completely closed and the pump (artificial heart-lung apparatus) was turned off, but the leaflet was stacked again, resulting in severe aortic regurgitation (ar).The pump was turned on and the valve was explanted interprocedurally.The leaflets are moving normally after the explant procedure.A new 23mm epic valve was implanted as replacement.There was some delay in the procedure.The patient was reported stable.
|
An event of physical resistance/sticking of the leaflets was reported and did not confirmed.Morphological and hydrodynamic examination indicated the valve met abbott specifications.Hydrodynamic testing upon return to abbott and at the time of manufacturing indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|