• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL G2 STRAIGHT JAW 14CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL G2 STRAIGHT JAW 14CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number NSLG2S14
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2024
Event Type  malfunction  
Event Description
It was reported that during an unknown procedure the ¿replace instrument¿ alert screen was displayed.Changed to another device to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Manufacturer Narrative
(b)(4) date sent: 2/21/2024 d4 batch #: v95n40 investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was returned with the upper jaw detached and not returned.In addition, the top of the iblade was broken and lifted.Due to the condition of the device returned we were unable to investigate further the issue reported.The device was connected to the generator and it was recognized.Because the jaw was detached not all functional testing could be performed with the generator.A probable cause of the damage could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.Please reference the instruction for use for more information: as part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.A manufacturing record evaluation was performed for the finished device batch v95n40, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL G2 STRAIGHT JAW 14CM
Type of Device
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18748627
MDR Text Key336648468
Report Number3005075853-2024-01486
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K112033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNSLG2S14
Device Lot NumberV95N9V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2024
Initial Date Manufacturer Received 02/06/2024
Initial Date FDA Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/25/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
GENERATOR
-
-