|
Model Number CDHFA500Q |
Device Problems
Failure to Capture (1081); High impedance (1291); Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/02/2024 |
Event Type
malfunction
|
Event Description
|
Related manufacturing reference number: 2017865-2024-33189 related manufacturing reference number: 2017865-2024-33191 it was reported that the left ventricular lead exhibited loss of capture.Dislodgement was confirmed via x-ray.The physician decided to revise the lead.When the physician attempted to reposition the lead, it was noted the guidewire was unable to advance through the lead.The lead was explanted and replaced.When the new lead was plugged into the chronic implantable cardioverter defibrillator (icd), it was noted there was loss of capture, high pacing impedance, and the set screw had stripped.The icd was removed and replaced.When the new icd was implanted, it was noted the set screw on the right ventricular header could not be removed.The icd was removed and replaced successfully.The patient was stable.
|
|
Manufacturer Narrative
|
The reported event of set screw anomaly was confirmed.Final analysis found the device was above the elective replacement indicator (eri) upon receipt.Analysis revealed the left ventricular septum was damaged during procedure and the set screw was unseated.The set screw anomaly was consistent with having occurred during the procedure.The reported event of capture and high pacing lead impedance could not be confirmed.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.During analysis, a failure event was observed which was unrelated to the reported event.Final analysis found that the header was partially opaque.
|
|
Search Alerts/Recalls
|
|
|