SMITH & NEPHEW, INC. JOURNEY DCF AP FEM CUT BLK 6; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number 74012416 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/31/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during a tka, while impacting the journey dcf ap fem cut blk 6, broke on femur.The attachment pins from the block broke from impaction.No pieces fell into patient.Surgery was performed, without any delay, with the same device.No patient complications were reported.
|
|
Manufacturer Narrative
|
H10: internal complaint reference:(b)(4).
|
|
Manufacturer Narrative
|
The associated device was returned and evaluated.The visual inspection revealed that the spiked cross bar detached from the body of the device as the cross bar screw seems to be missing.The device shows signs of extensive wear and use.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed a similar event for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|