C.R. BARD, INC. (BASD) -3006260740 BD PREVUE II PERIPHERAL VASCULAR ACCESS SYSTEM 20MM TRADITIONAL PROBE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
Catalog Number 9770703 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/26/2024 |
Event Type
malfunction
|
Event Description
|
Found during evaluation at bd service center: the ultrasound image produced is very dark.The transducer element test shows 6 bad elements out of 8 total.The root cause is due to an internal failure of the probe.
|
|
Manufacturer Narrative
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The site rite prevue ii cue probe was visually inspected upon receipt and was found to be in poor physical condition.The reported issue of probe doesn't work is confirmed.The ultrasound image produced is very dark.The transducer element test shows 6 bad elements out of 8 total.The root cause is due to an internal failure of the probe.
|
|