Catalog Number CDS0705-NTW
Device Problem Incomplete Coaptation (2507)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2024, a patient presented with grade 4 functional mitral regurgitation (mr) for a mitraclip procedure.One clip was placed on anterior 2 and posterior 2 leaflet segments (a2-p2).It was noted that imaging was challenging depicting the echocardiogram images.The patient was stable after deployment.On (b)(6) 2024, the patient returned for a transthoracic echocardiogram and a single leaflet device attachment was noted.The mr severity is unknown, but it was decreased from pre-procedure mr.The recurrent mr was stabilized with medication to grade 1-2.A second intervention will not be performed due to minimal grasping area and the age of the patient.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.All available information was investigated and the reported slda per the account is related to procedural conditions associated with difficulty depicting the echocardiogram images.Image resolution poor is related to patient conditions associated with difficult imaging.Mitral valve insufficiency/ regurgitation appears to be due to the slda.Mitral regurgitation is listed in the instructions for use (ifu) as a known possible complication associated with mitraclip procedures.Medication required was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18760240
MDR Text Key336966225
Report Number2135147-2024-00829
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/12/2024
Device Catalogue NumberCDS0705-NTW
Device Lot Number30314R1108
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/01/2024
Initial Date FDA Received02/22/2024
Supplement Dates Manufacturer Received02/22/2024
Supplement Dates FDA Received02/27/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age74 YR
Patient SexMale
-
-