Device Problem
Device Dislodged or Dislocated (2923)
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type
malfunction
Event Description
It was reported that during the initial implant procedure, dislodgment of the left ventricular (lv) lead occurred.The physician suspected the lv lead dislodgment was due to the anatomy of the patient.The lv lead was explanted and replaced to resolve the event.The patient was in stable condition.
Manufacturer Narrative
The reported event was lead dislodgement.As received, a complete lead was returned in one piece for analysis.Visual inspection of the lead found no anomalies with the exception of damage consistent with procedural damage.Electrical testing did not find any indication of conductor fractures or internal shorts.The s-curve hump height was measured within specifications.