Trackwise#: (b)(4) updated sections: b4, g3, g6, h2, h3, h6, h10 the device was returned to the factory for evaluation on 02/13/2024.An investigation was conducted on (b)(6) 2024.A visual inspection was conducted.Signs of clinical use and evidence of charred blood was observed on the heater wire.There were no visual defects observed on the intact clear silicone insulation on both the cold and hot jaws.The heater wire was observed to be flexed away from the hot jaw from the base of the hot jaw with detachment at the tip of the hot jaw.No other visual defects were observed.An electrical evaluation was conducted.A pre-cautery test was performed per the direction for use (dfu) with a reference cable, adapter, and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it produced visible steam and heat during ten (10) 3-second activations and shut off when the toggle was released.No excessive smoke and/or steam were observed during the testing.To evaluate the safety shut down system, a polyfuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.An activation and transection capability test was performed over four (4) repetitions using "max life test method stm2048073 rev aa.The device successfully transected tissue four (4) times.No final testing was conducted due to the conditon of the heater wire.Based on the returned condition of the device, and the no specific failure reported, the analyzed failure "material twisted/bent wire " was observed.The lot # 3000350756 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|