• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number BELLAVISTA 1000E US 17,3 VENTILATOR
Device Problems Defective Alarm (1014); High Readings (2459)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2024
Event Type  malfunction  
Event Description
The customer reported to vyaire medical problem with bellavista 1000e us that would not stop alarming.As per the alarm logs, technical failure 274 (flow measured proximally is greater than the flow delivered) and tf 105 (volume low) were found.Furthermore, there was no information regarding patient associated with the event.
 
Manufacturer Narrative
Vyaire medical file identification: (b)(4).H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BELLAVISTA
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
IMTMEDICAL AG
gewerbestrasse 8
buchs sankt gallen 9470
SZ  9470
Manufacturer (Section G)
TECHNOCOM SYSTEMS SDN BHD
plo 1, jalan firma 1, kawasan
perindustrian tebrau 1
johor bahru 81100
MY   81100
Manufacturer Contact
sandra valencia
510 technology drive
irvine, CA 92618
8477865998
MDR Report Key18763407
MDR Text Key336970456
Report Number3004553423-2024-02182
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBELLAVISTA 1000E US 17,3 VENTILATOR
Device Catalogue Number301.100.130
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/29/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-