Model Number BELLAVISTA 1000E US 17,3 VENTILATOR |
Device Problems
Defective Alarm (1014); High Readings (2459)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/29/2024 |
Event Type
malfunction
|
Event Description
|
The customer reported to vyaire medical problem with bellavista 1000e us that would not stop alarming.As per the alarm logs, technical failure 274 (flow measured proximally is greater than the flow delivered) and tf 105 (volume low) were found.Furthermore, there was no information regarding patient associated with the event.
|
|
Manufacturer Narrative
|
Vyaire medical file identification: (b)(4).H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|