• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; IMPLANTABLE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; IMPLANTABLE DEVICE Back to Search Results
Model Number G148
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2024
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed that the power consumption is increasing in a gradual fashion.This appears consistent with capacitor degradation due to hydrogen gas content in the sealed pulse generator (pg) atmosphere.Furthermore, a device replacement was recommended or reassessed battery status within 90 days.To date, this device remains in service.No adverse patient effects were reported.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed that the power consumption is increasing in a gradual fashion.This appears consistent with capacitor degradation due to hydrogen gas content in the sealed pulse generator (pg) atmosphere.Furthermore, a device replacement was recommended or reassessed battery status within 90 days.To date, this device remains in service.No adverse patient effects were reported.Additional information indicated that this device was explanted and replaced.No additional adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18764536
MDR Text Key336203626
Report Number2124215-2024-10761
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/02/2017
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number136369
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/01/2024
Initial Date FDA Received02/22/2024
Supplement Dates Manufacturer Received03/25/2024
Supplement Dates FDA Received04/07/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
-
-