MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105NNGYNA NOVO NORDISK GRAY; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105NNGYNA NOVO NORDISK GRAY; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNGYNA
Device Problems Computer Software Problem (1112); Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the customer noticed inpen dose(s) are not transmitted to inpen app.It was also reported that inpen is not functioning and no longer auto dosing have to enter the doses manually.Troubleshooting was performed and the issue was not resolved.No harm requiring medical intervention was reported.The customer will discontinue the use of the device.The device will be returned for analysis.
 
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Per visual inspection: no physical damage to cartridge holder or inpen front and back shell was noted.Several attempts were made to pair inpen, every time app displayed inpen not found".The inpen does not pair with commercial mobile app.Inpen did not transmit to manufacturing app.Unable to perform baseline wireless/functionality test or displacement doe accuracy test.Inpen passed dialing alignment using dose knob zero alignment gage.Performed battery investigation and measured.0349 v.Unit reached end of life.Unable to confirm dose log inaccuracy due to inpen not communicating to mobile app.In conclusion: inpen failed to communicate to commercial app or manufacturing app.Inpen battery lifetime last 1 year after first paring.It is possible battery life decreased as expected and the low battery icon appears several times near the end of the battery lifetime.Inpen working as designed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNGYNA NOVO NORDISK GRAY
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
lusine boyadzhyan
1223 world trade drive, 100
san diego, CA 92128
MDR Report Key18764849
MDR Text Key336523061
Report Number3012822846-2024-00262
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/23/2023
Device Model NumberMMT-105NNGYNA
Device Catalogue NumberMMT-105NNGYNA
Device Lot NumberB0531
Was Device Available for Evaluation? No
Date Manufacturer Received02/07/2024
Date Device Manufactured08/23/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age71 YR
Patient SexFemale
Patient Weight51 KG
-
-