Model Number 1456Q/86
Device Problem Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/05/2024
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the left ventricular lead exhibited a high capture threshold.The left ventricular lead was not used and replaced.The patient was in stable condition throughout the procedure.
 
Manufacturer Narrative
The reported event of inadequate threshold was not confirmed.As received, a complete lead was returned in one piece for evaluation.Electrical tests did not find any indication of conductor fractures.Visual and x-ray inspections of the lead did not find any anomalies.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications before leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18765058
MDR Text Key336318489
Report Number2017865-2024-33326
Device Sequence Number1
Product Code NIK
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1456Q/86
Device Lot NumberA000145646
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-