Evaluation of the returned ruby coil pusher assembly confirmed that the pusher assembly was fractured, and the embolization coil was detached.If the ruby coil is manipulated at an angle during preparation, the pusher can become kinked.This damage likely contributed to the reported resistance experienced during advancement of the ruby coil.Subsequent use of a previously damaged pusher and additional manipulation against resistance likely lead to the pusher assembly fracturing.If the pusher assembly fractures, the embolization coil will likely detach.The detached embolization coil was not returned for evaluation.Further evaluation of the device revealed additional kinks along the length of the pusher assembly.Based on the returned packaging condition, this damage was incidental to the complaint and likely occurred during packaging for the device return.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a coil embolization procedure in a branch of the gastroduodenal artery (gda) using ruby coils and a non-penumbra microcatheter.During the procedure, the nurse kinked the ruby coil pusher wire and then attempted to straighten it.The physician then experienced resistance while advancing the ruby coil into the microcatheter and subsequently noticed that the pusher wire was fractured.While removing the distal end of the ruby coil pusher wire, the ruby coil unintentionally detached in the proximal end of the microcatheter.The physician then manually aspirated the ruby coil out of the microcatheter using a syringe.The procedure was completed using additional ruby coils and the same microcatheter.There was no report of an adverse effect to the patient.
|