It was reported that during an arthroscopy, the needle of the firstpass suture passer was detached from the device after passing the ultrabraid suture through the rotator cuff.Did a search from the patient drape and floor, even did a x-ray on the patient to ensure it was not inside the patient shoulder.However the needle could not be found.The procedure was completed using a back-up device.There was a delay less than 30 minutes and no further complications were reported.
|
The reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual evaluation showed the device was not returned in original packaging.Bio debris is present.There is no other visible defect.A functional evaluation showed pulling the lever will close and lock the jaw.Pulling the lever and trigger simultaneously will not deploy the suture passer needle.A destructive evaluation performed by the complaints engineer, of opening the device, verified the needle is inside the device.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, attempted correction of a damaged device, off-axis insertion, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|