An adverse skin reaction was reported with wear of the adc device.Customer experienced symptoms described as "pain, pus coming out of wound, bump the size of golf ball below skin, and nausea" and had contact with a healthcare professional who provided an unspecified topical ointment for treatment.There was no report of death or permanent impairment associated with this event.
|
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned and a valid serial number has not been provided.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor products.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.Clinical data was reviewed and confirmed that freestyle libre sensors continue to be safe, effective, and perform as intended in the field.A tripped trend review was conducted for the reported complaint and freestyle libre sensors, no trends were identified that would indicate any product related issues.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|