• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL EU 4.5X22MM STENT 12 MM DW TIP; INTRACRANIAL NEUROVASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL EU 4.5X22MM STENT 12 MM DW TIP; INTRACRANIAL NEUROVASCULAR STENT Back to Search Results
Catalog Number ENC452212
Device Problems Premature Activation (1484); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  malfunction  
Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Section e.1: the initial reporter phone: (b)(6).The initial reporter email address was not available / reported.Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.(b)(6) medical performed a review of the device history records relative to the manufacturing, inspection, and packaging of the lot 8294528.The history record indicates this product was final inspection tested at (b)(6) medical and was determined to be acceptable.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event Description
The healthcare professional reported that during an endovascular embolization procedure, the 4.5mm x 22mm enterprise® vascular reconstruction device (enc452212 / (b)(6)) encountered some resistance in the y-connector.The physician retracted the stent and it was already automatically released; the stent component became prematurely separated from the delivery wire.The physician replaced the stent to complete the procedure using the same microcatheter.There was no report of any negative patient impact.On 22-feb-2024, additional information was received.Per the information, the target vessel of the procedure was the m1 segment of the middle cerebral artery (mca).A continuous flush had been maintained through the microcatheter.The microcatheter used was recantol (unknown manufacturer).The stent / stent delivery system did not appear damaged.Nothing unusual was noted about the system prior to use.The replacement stent was a 4.5mm x 28mm enterprise® vascular reconstruction device (enc452812) there was no clinically significant delay in the procedure as a result of the reported issue.
 
Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product analysis lab received the complaint device on 11-mar-2024.A supplemental 3500a report will be submitted once the product investigation has been completed.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report the investigational finding of the returned device.The complaint product was returned and received for evaluation and analysis.The investigation is documented below.Investigation summary: a non-sterile 4.5mm x 22mm enterprise® vascular reconstruction device was received contained in the decontamination pouch.Visual inspection was performed.It was noted that the stent was already detached from the unit.The delivery wire and the introducer were in good condition (i.E., no kinks, no bends, nor elongations).Microscopic inspection was performed on the stent component.It was observed to be in good condition; there was no structural damage (i.E., no broken struts, no kinks); also, it was noted fully expanded, both ends can be noted as completely flared.Dimensional analysis was performed on the delivery wire.All measurements were found to be within specifications, including the specifications that control the attachment and delivery of the stent.Therefore, device failure is not suspected to be a contributing factor.The reported issue documented in the complaint regarding the stent automatically released was confirmed since the stent was noted as already separated from the delivery system; based on this condition, the issue regarding some resistance in the y-connector cannot be evaluated through functional testing.Additionally, none of the returned components present damages that suggest that they were forcibly advanced.With the limited information available, a conclusive cause cannot be determined; however, it is possible that clinical and procedural factors, including device manipulation and operator's technique, may have contributed to the reported failure.At this time, there is no evidence to support that the issue reported in the complaint is a result of a defect inherently related to the device.Lake region medical performed a review of the device history records relative to the manufacturing, inspection, and packaging of the lot 8294528.The history record indicates this product was final inspection tested at lake region medical and was determined to be acceptable.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) does contain the following recommendations: ¿ if resistance is met during manipulation, determine the cause of resistance before proceeding.¿ do not apply undue force if resistance is encountered at any point during stent manipulation.Withdraw the unit and advance a new one.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EU 4.5X22MM STENT 12 MM DW TIP
Type of Device
INTRACRANIAL NEUROVASCULAR STENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS, INC.
325 paramount dr
raynham MA 02767
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-689
MDR Report Key18772942
MDR Text Key337075234
Report Number3008114965-2024-00214
Device Sequence Number1
Product Code NJE
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
H60001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberENC452212
Device Lot Number8294528
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2024
Initial Date FDA Received02/23/2024
Supplement Dates Manufacturer Received03/11/2024
03/12/2024
Supplement Dates FDA Received03/11/2024
03/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
RECANTOL MICROCATHETER
-
-