The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.A visual inspection and screening test of the returned device were performed in accordance with bwi procedures.Visual analysis revealed holes in the pebax and foreign reddish material was observed inside it.The root cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.A screening test was performed, and the device was visualized and recognized correctly; however, the force vector was observed inverted and high force was observed, since the sensor values were found within specifications, this issue could be related to the damage on the pebax.A manufacturing record evaluation was performed for the finished device 31180249l, and no internal action was found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use contain the following recommendations: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4);.
|
It was reported that a patient underwent an atrial fibrillation ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, the force values would jump to high force when ablation happened.When the catheter was taken out of the body the medical team discovered discoloration on the sides.The medical team believed that the discoloration was possible blood.When the catheter was replaced, the issue was resolved.No patient consequences were reported.
|