• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB COCHLEAR¿ VISTAFIX® VXA300 ABUTMENT 4.5MM; COCHLEAR BAHA VISTAFIX SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB COCHLEAR¿ VISTAFIX® VXA300 ABUTMENT 4.5MM; COCHLEAR BAHA VISTAFIX SYSTEM Back to Search Results
Model Number 92995
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Skin Infection (4544)
Event Type  Injury  
Event Description
Per the clinic, the patient experienced an infection at implant site and subsequently was treated with oral antibiotics (specific date and duration not reported).Additional information has been requested but has not been made available as of the date of this report.
 
Manufacturer Narrative
This report is submitted on february 26, 2024.
 
Manufacturer Narrative
Per the clinic, the patient was placed under general anesthesia on (b)(6) 2024 in order to undermine the skin and remove the abutment.The patient was also administered with topical antibiotics (specific date and duration not reported).The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.This report is submitted on may 09, 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCHLEAR¿ VISTAFIX® VXA300 ABUTMENT 4.5MM
Type of Device
COCHLEAR BAHA VISTAFIX SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
asyikin zulkhairi
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key18775396
MDR Text Key336176107
Report Number6000034-2024-00565
Device Sequence Number1
Product Code FZE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K945154
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number92995
Device Catalogue Number92995
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/30/2024
Initial Date FDA Received02/26/2024
Supplement Dates Manufacturer Received04/12/2024
Supplement Dates FDA Received05/09/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-