Model Number 71953-01
Device Problem Loss of Data (2903)
Patient Problems Hyperglycemia (1905); Loss of consciousness (2418)
Event Date 01/28/2024
Event Type  Injury  
Manufacturer Narrative
The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the freestyle libre reader were reviewed and the dhrs showed the freestyle libre reader passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported a "missing or incorrect data in memory" issue with the abbott diabetes care (adc) device.As a result, the customer experienced a loss of consciousness and received treatment of rapid insulin provided by a non-healthcare professional.No further treatment details were provided.There was no report of death or permanent impairment associated with this event.
 
Event Description
A customer reported a "missing or incorrect data in memory" issue with the abbott diabetes care (adc) device.As a result, the customer experienced a loss of consciousness and received treatment of rapid insulin provided by a non-healthcare professional.No further treatment details were provided.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Reader (b)(6) has been returned and investigated.Visual inspection has been performed on the returned reader and minor damage usb port was observed, however the port functions as intended.Sensor data was analyzed.Graph was missing due to customer did not scan within specific time frame of the previous sensor scan.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18776493
MDR Text Key336185897
Report Number2954323-2024-06663
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71953-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 02/13/2024
Initial Date FDA Received02/26/2024
Supplement Dates Manufacturer Received04/07/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-