Model Number G148 |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the battery of this cardiac resynchronization therapy defibrillator (crt-d) went from 10 months to four months of battery remaining.Boston scientific technical services (ts) explained the elective replacement indicator (eri), and small changes in output or pacing percentage can change the projected longevity significantly as the battery nears explant.To date, the device remains in service.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that the battery of this cardiac resynchronization therapy defibrillator (crt-d) went from 10 months to four months of battery remaining.Boston scientific technical services (ts) explained the elective replacement indicator (eri), and small changes in output or pacing percentage can change the projected longevity significantly as the battery nears explant.Subsequently, this crt-d was explanted and replaced.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|