Catalog Number 620010001
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Event Description
It was reported that the version indicator is cracked.Patient status/ outcome / consequences: no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study: unknown, ip-01978775.Device property of: none, device in possession of: none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
 
Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : according to the information received, ¿cracked¿.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the version indicator has broken at one of the distal prongs, fragments was not returned for evaluation.The observed condition of the device was consistent with a component failure that may have been caused by exposure to unintended forces, prying motion when disassembling the device is suspected.The overall complaint was confirmed as the observed condition of the version indicator would contribute to the a device issue.Based on the investigation findings, potential cause can be attributed to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSION INDICATOR
Type of Device
ALIGNMENT DEVICE
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18777950
MDR Text Key336213883
Report Number1818910-2024-04307
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number620010001
Device Lot Number321000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-