The reported event of inadequate high voltage (hv) output and premature battery depletion could not be confirmed.The implantable cardioverter defibrillator (icd) was returned for analysis.Visual inspection of the device did not reveal any anomaly that could contribute to the reported event.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.A longevity calculation was performed and found the battery depletion was normal based on the device usage.
|