A high readings issue with the adc device was reported.Customer received unspecified higher sensor scan results compared to unspecified blood glucose readings obtained on a competitor brand meter and experienced symptoms of hypoglycemia, sweating, and confusion.The customer was provided with milk and biscuits for treatment by a non-healthcare professional for treatment.There was no report of death or permanent impairment associated with this event.
|
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
Sensor (b)(6)has been returned and investigated.The sensor plug is fully seated and no issues were observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 6 (indicating early termination).Visual inspection has been performed on the sensor plug assembly.Unable to reprogram the sensor.Sensor was de-cased and battery was replaced.The current was applied to the sensor to perform accuracy testing while in the test fixture.All results were within specification.Poise voltage and sensor thermistor testing were both within specification, indicating the sensor was providing accurate glucose readings.Therefore, issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
A high readings issue with the adc device was reported.Customer received unspecified higher sensor scan results compared to unspecified blood glucose readings obtained on a competitor brand meter and experienced symptoms of hypoglycemia, sweating, and confusion.The customer was provided with milk and biscuits for treatment by a non-healthcare professional for treatment.There was no report of death or permanent impairment associated with this event.
|