This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation and device evaluation.The device was evaluated by olympus.And it was confirmed, that the ceramic tip is broken.Additionally, there were non-reportable (non-pae) defects noted.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely, the reported event occurred, due to wear and tear, wrong handling or cracks in the insulation material.Which are not visible in most cases.Additionally, it is possibly caused by the impact of a major mechanical force, e.G.A blow or a fall.Based on the nature of the damage, it is assumed, that this is a thermo-mechanically induced damage.It is unclear, whether the insulation insert had a previous damage or wear and tear through reprocessing or from the last procedure.Furthermore, fragments of the ceramic isolation inlet will usually be removed by the surgeon, during the procedure.If the fragments are not found, they can be located using x-ray or ct scanning and are subsequently removed.The root cause of the reported event could not be determined.The event can be detected/prevented, by following the instructions for use, which state: warning infection control risk: "properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel".4.1: inspection and testing: inspecting the product: "visually inspect the product.Make sure that it has: no corrosion, no dents, no scratches.Ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures)".Warning risk of injury: "impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user"."do not use the instrument if damaged".Olympus will continue to monitor field performance for this device.
|