The device contained 100ml of dextrose 5% in water.The lot was manufactured between august 10, 2023 - august 14, 2023.The actual device was received for evaluation.Visual inspection was performed which identified fluid inside a bag that contained the unit.The cause of leak inside the bag was found to be an untightened winged luer cap.Signs of surface roughness were not observed inside the cap when inspected under the microscope.A functional leak test was performed by filling the unit with green color water and the cap was hand tightened; the device was monitored till the next day and no signs of leak were observed.The device was determined to be conforming product during functional testing.The reported condition was verified during inspection; however, not verified during functional testing.The cause of the condition could not be determined; however, the probable cause is use-related due to not securely tightening the winged luer cap.The product label (ifu, instructions for use) indicates to ensure the cap is securely connected after filling and priming.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|