DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-52 |
Device Problem
Break (1069)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/20/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) top cover cracked.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during a check, the cardiosave intra-aortic balloon pump (iabp) top cover cracked.There was no patient involvement.
|
|
Manufacturer Narrative
|
Corrected data: b5, b6, b7, d10, h6 (clinical and clinical code).Updated data: b4, g3, g6, h2, h10, h11.
|
|
Manufacturer Narrative
|
Additional information: event postal code: (b)(6).A getinge field service engineer (fse) evaluated and found slight crack on top cover of screen.Fse replaced display top cover assy with tape kit.The unit passed all functional and safety tests to factory specifications.Unit was cleared for clinical use.
|
|