Model Number HL-390
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Event Description
It was reported that the unit was leaking.The malfunction was found during preventive maintenance.No patient harm/adverse event reported.
 
Manufacturer Narrative
B3: date of event and d5: operator of device are unknown, no information has been provided to date.H3 - other: device has not been returned to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
H6 - updated.Device evaluation: no product was returned for investigation.Therefore, we are unable to determine the reported complaint.If we receive the device, we will reopen the investigation for further evaluation.However, based on the reported problem, the most probable cause is a cracked tank cover and/or a cracked tank by the drain tube fitting.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18791478
MDR Text Key336364434
Report Number3012307300-2024-01019
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/19/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-