• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. WEB SL; INTRASACCULAR FLOW DISRUPTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. WEB SL; INTRASACCULAR FLOW DISRUPTION DEVICE Back to Search Results
Model Number W5-3.5-2-MVI
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2023
Event Type  malfunction  
Manufacturer Narrative
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Investigation conclusion: the investigation of the returned web system found the hypotube kinked, the proximal connector kinked, and the heater coil windings stretched.The microcatheter used with the web system in the procedure was found kinked at 1cm and 3cm from the distal tip.The web device failed continuity and resistance testing due to the damaged connector and heater coil windings, which is consistent with the alleged detachment issue.However, the heater coil pet and implant tether were found melted, indicating that the device was activated using a detachment controller during the procedure; therefore, the damage to the connector and heater coil windings likely occurred post-activation.The stretched heater coil windings are an indication that the tether could have been caught in between the coil windings and caused the heater coil to stretch when the delivery system was pulled/retracted during the procedure.The physical evaluation of the device could not identify the conditions or circumstances that led to the pusher damage, but the damage is consistent with the device experiencing forces over specification.The returned detachment controllers were found to function as intended and would not have caused or contributed to the reported event.Potential complications as referenced on the ifu, include but are not limited to the following: hematoma at the site of entry, aneurysm rupture, emboli, vessel perforation, parent artery occlusion, hemorrhage, ischemia, vasospasm, clot formation, device migration or misplacement, premature or difficult device detachment, non-detachment, incomplete aneurysm filling, revascularization, post-embolization syndrome, and neurological deficits including stroke and death.
 
Event Description
A media ¿ aneurysm treatment was performed with a web device.Previous connectivity check with detachment controller was successful; however, after the web was introduced and placed inside the aneurysm, detachment of the device was not successful.The physician attempted to detach the web several times and switched to another detachment controller without success.The web was removed from the patient and re-sheathed, and the aneurysm was successfully treated with coils.The patient was reported to be in good condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEB SL
Type of Device
INTRASACCULAR FLOW DISRUPTION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key18794610
MDR Text Key336517836
Report Number2032493-2024-00170
Device Sequence Number1
Product Code OPR
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P170032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberW5-3.5-2-MVI
Device Lot Number0000348356
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
WDC-2 DETACHMENT CONTROLLER
-
-