A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Investigation conclusion: the investigation of the returned web system found the hypotube kinked, the proximal connector kinked, and the heater coil windings stretched.The microcatheter used with the web system in the procedure was found kinked at 1cm and 3cm from the distal tip.The web device failed continuity and resistance testing due to the damaged connector and heater coil windings, which is consistent with the alleged detachment issue.However, the heater coil pet and implant tether were found melted, indicating that the device was activated using a detachment controller during the procedure; therefore, the damage to the connector and heater coil windings likely occurred post-activation.The stretched heater coil windings are an indication that the tether could have been caught in between the coil windings and caused the heater coil to stretch when the delivery system was pulled/retracted during the procedure.The physical evaluation of the device could not identify the conditions or circumstances that led to the pusher damage, but the damage is consistent with the device experiencing forces over specification.The returned detachment controllers were found to function as intended and would not have caused or contributed to the reported event.Potential complications as referenced on the ifu, include but are not limited to the following: hematoma at the site of entry, aneurysm rupture, emboli, vessel perforation, parent artery occlusion, hemorrhage, ischemia, vasospasm, clot formation, device migration or misplacement, premature or difficult device detachment, non-detachment, incomplete aneurysm filling, revascularization, post-embolization syndrome, and neurological deficits including stroke and death.
|
A media ¿ aneurysm treatment was performed with a web device.Previous connectivity check with detachment controller was successful; however, after the web was introduced and placed inside the aneurysm, detachment of the device was not successful.The physician attempted to detach the web several times and switched to another detachment controller without success.The web was removed from the patient and re-sheathed, and the aneurysm was successfully treated with coils.The patient was reported to be in good condition.
|