The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd module with drive pcb.In addition, our technician confirmed that the segment fluid damage, the ccd drive pcb fluid damage, the electrical pin connector fluid damage, the lg connector fluid damage, the lcb (light carrying bundle) broken, the remote control buttons perforated, the insertion flexible tube buckled, and the u/d and the r/l pulley wires worn out; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|