Model Number 1458QL/86
Device Problems Failure to Capture (1081); Over-Sensing (1438); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  Injury  
Event Description
It was reported that the patient presented remotely via merlin.Net.Upon review of the transmission, it was noted that the left ventricular lead exhibited dislodgement, loss of capture and oversensing issue.Fluoroscopy was performed and the left ventricular lead was found to be dislodged.The left ventricular lead was explanted and replaced.The patient was in stable condition throughout the procedure.
 
Manufacturer Narrative
The reported events were dislodgement, failure to capture, and over-sensing.As received, a complete lead was returned in one piece for analysis.The reported events of failure to capture and over-sensing were not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.A visual and x-ray inspection found no anomalies, except for procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18795392
MDR Text Key336393954
Report Number2017865-2024-33653
Device Sequence Number1
Product Code NIK
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458QL/86
Device Lot NumberA000143661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA STS OPTIM; GALLANT HF ICD; TENDRIL STS
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-