Model Number MNA-140 PCIE KS |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Barco has concluded that the loss of video signal is caused by a problem with the sfp modules, which is related to maintenance.After replacement of the sfp modules the problem was solved.We have no feedback since then despite several attempts.
|
|
Event Description
|
The customer has informed barco that they received the following information from the hospital: image loss intraoperatively.There will now be a swissmedic message from the customer.Intraoperative image failure of the two monitors (dec dp4k) in the sterile field and the 55" barco with built-in dec on the wall.Only the secureline would have worked (12g-sdi) but was switched incorrectly.Now the customer has again received the setting for switching the secureline via video.I played it through with him via video support.It works perfectly.As far as i know, no patient was harmed.
|
|