• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARCO NV BARCO; NEXXISOR, MNA-440 ENC SDI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARCO NV BARCO; NEXXISOR, MNA-440 ENC SDI Back to Search Results
Model Number NEXXISOR, MNA-440 ENC SDI
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2022
Event Type  malfunction  
Event Description
Reported failure: after the introduction of the nis for the interor setup, the black screen issue stabilized for most rooms, except for 2 rooms.+ issue determination difficult: occurrence 1x per month.+ ntp server introduced for better timing determination.+ logfiles provided by support.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARCO
Type of Device
NEXXISOR, MNA-440 ENC SDI
Manufacturer (Section D)
BARCO NV
president kennedypark 35
kortrijk, west-vlaanderen 8500
BE  8500
Manufacturer Contact
lieven de wandel
president kennedypark 35
kortrijk, west-vlaanderen 8500
BE   8500
MDR Report Key18796477
MDR Text Key336516006
Report Number3004455125-2024-00010
Device Sequence Number1
Product Code DXJ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K173381
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Remedial Action Modification/Adjustment
Type of Report Initial
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberNEXXISOR, MNA-440 ENC SDI
Device Catalogue NumberK9303301
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/23/2022
Initial Date FDA Received02/28/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-