Model Number MNA-240 ENC DP KS |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A medical specialist has confirmed that the extension of anesthesia, as reported by the customer, did not pose an additional risk to the patient.
|
|
Event Description
|
The customer has informed barco that they received the following information from the hospital: all monitors are black when booting up in the morning.System up and down several times.Time delay approx.20 minutes until the decoder on the equipment trolley worked and sent a signal.Everything was stable for the rest of the day.The next morning there were no problems at starting.It was a user error that the emergency line was on the wrong video input /4k instead of 1080p.It is not clear who set this.The patient received extended anesthesia.The surgery could not be started.
|
|