• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; NO MATCH Back to Search Results
Model Number CDVRA500Q
Device Problems Premature Discharge of Battery (1057); Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that distributed implantable cardioverter defibrillator was found to have premature battery depletion and charge number exceeded the limit during stock rotation verification.
 
Manufacturer Narrative
Device was returned from stock rotation and analyzed in the lab.The stock rotation reported event of excessive charge and battery consumption was confirmed.However, the cause of the reported event was determined to be due to the device having been taken out of ship settings in the field.The device functioned as expected and according to its programmed settings.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18797297
MDR Text Key336888775
Report Number2017865-2024-33659
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Model NumberCDVRA500Q
Device Lot NumberS000085571
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/09/2024
Initial Date FDA Received02/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-