• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION STOPCOCK; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION STOPCOCK; STOPCOCK, I.V. SET Back to Search Results
Catalog Number ZMC7401
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2024
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter facility name : (b)(6) hospital.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a stopcock in blister pack was cracked which resulted in a leak.The event occurred when attempting to connect the stopcock to an iv infusion extension line.The stopcock was replaced.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was not available; however, three (3) retained samples were evaluated.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STOPCOCK
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
marsa
MT  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18802015
MDR Text Key336519028
Report Number1416980-2024-00824
Device Sequence Number1
Product Code FMG
UDI-Device Identifier05413760157153
UDI-Public(01)05413760157153
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberZMC7401
Device Lot Number23A06V134
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/02/2024
Initial Date FDA Received02/28/2024
Supplement Dates Manufacturer Received03/01/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-