• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2024
Event Type  malfunction  
Event Description
During the initial implant procedure, it was not possible to remove the stylet from the left ventricular (lv) lead.The lv lead was explanted and a follow-up procedure is planned to attempt lv lead replacement.The patient was stable with no consequences.
 
Manufacturer Narrative
As received, a complete lead with stuck stylet was returned in one piece.The polytetrafluoroethylene (ptfe) coating of the stuck stylet was stripped and was found bunched up with the inner coil distal to the connector pin.The connector pin with the cap and crimp sleeve were found pulled out of the connector assembly consistent with damage due to excessive forces applied while attempting to remove the stylet from the lead during the procedure.The cause of the reported event was isolated to the bunching of the stylet ptfe coating inside the inner coil at the connector region that prevented the removal of the stylet and excessive forces resulted in the connector pin to be pulled out of the connector assembly.Electrical testing did not find any indication of conductor fractures or internal shorts.A review of the device history record (dhr) confirmed that no issues were identified related to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18806530
MDR Text Key336545754
Report Number2017865-2024-22652
Device Sequence Number1
Product Code NIK
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberS000091931
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/13/2024
Initial Date FDA Received02/29/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; ENTRANT HF; TENDRIL STS
-
-