Reportable based on device analysis completed on 09feb2024.It was reported that the coil could not be pushed in the middle of the angiography catheter.The target lesion was located in the gastric fundus vein.A 10mm x 30cm interlock was selected for use.During the procedure, the coil was advanced into stc microcatheter and it was found out that the coil could not be pushed in the middle of angiography catheter.Re-flushing with heparin saline was done but same issue occurred.The coil was withdrawn, and the procedure was completed with another of the same device.No complications were reported, and patient was stable post procedure.However, device analysis revealed that the main coil was detached.
|