Model Number SC-1160
Device Problems High impedance (1291); Unexpected Therapeutic Results (1631); Delayed Charge Time (2586); Positioning Problem (3009)
Patient Problem Pain (1994)
Event Date 06/29/2020
Event Type  Injury  
Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Additional suspect medical device components involved in the event: product family: scs-linear leads upn: m365sc2317500 model: sc-2317-50 serial: (b)(6).Batch: 5099544/5086073.
 
Event Description
It was reported that the patient had inadequate pain relief and increased ipg charging time.High impedances were also noted on the leads.Database analysis showed signs of poor charger to ipg coupling and charger thermistor triggering often which could delay charging times.The impedance measurements revealed high values on nine contacts.The patient underwent a revision procedure wherein the ipg and leads were replaced with an mri compatible device.The explanted device components were kept by the medical facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA WAVEWRITER
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
*  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key18809818
MDR Text Key336582869
Report Number3006630150-2024-01081
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/11/2021
Device Model NumberSC-1160
Device Catalogue NumberSC-1160
Device Lot Number354935
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2024
Initial Date FDA Received02/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-